all EM beam machines produce x-rays because of beam-specimen interactions.
They come from the entire monte carlo shell of beam interaction--the
x-rays given off are photons and less likely than electrons to lose their
energy traveling through the sample. Typically, resolution of adjacent
element locations is much poorer compared to spot resolution. For example,
spatial resolution might be 4 nm, but the elemental resolution might be
30 um in a biological sample
Each x-ray produced is called a line and has a characteristic energy and wavelength.
Each x-ray comes from an electron shell vacancy being filled by another electron.
K, L, M, N
the subscripts denote how many shells the filling electron had to jump
down
e.g., a K shell vacancy filled by a M shell electron ==> Kb X-ray.When electrons jump down, they leave a hole behind them that will also need to be filled.
X rays will only be produced if the acc. voltage is high enough to knock out an electron
==>critical excitation energy or absorption edge ---this is the minimum amount of energy required to knock an electron from a given shell in a given element.
These values are known and can be looked up in the tables--use 1.5-3X the critical energy for a particular line and to insure a sufficient number of x-rays are produced (to be detectable)
K lines are higher energy x-rays than L lines. You need to dump more energy to jump into the K shell than the L shell.
Element |
K line absorption edge/minimum kV |
L line absorption edge/minimum kV |
Si |
1.7/2.55 kV |
--- |
Fe |
6.4/9.6 kV |
0.705/ 1.1 kV |
Pd |
21.2/31.8 kV |
2.84/4.26 kV |
Different elements have different excitation energies: the penetration
of the beam into the sample leads to dissipation of the beam. Elements
with lower critical excitation energies can be excited from deeper within
the monte carlo shell. This leads to differences in spatial resolution
and is more important in whole samples compared to thin sections.
In addition to the characteristic X rays, we see background, continuum or bremsstrahlung X-rays (different words for the same thing).
These arise as a result of the beam interacting with the nucleus and tightly
bound
electrons. See peaks superimposed over the background
X-rays have both a wavelength and an energy, and detectors can collect information on either
==>WDS wavelength dispersive spectrometer
the detector must be tuned for a given wavelengthworks best at higher energy--greater sensitivity
==>EDS energy dispersive spectrometer
cheaper detector--simpler design, robusthigh efficiency
can measure entire energy spectrum
use smaller beam size, currents
advantage of WDS==> spectral sensitivity-- the size of the window
of detection for each detector
The WDS can see a window 10eV wide vs. 135 eV wide for EDSIf peaks overlap from various elements, can't resolve them as easily with the EDS compared to WDS
EDS DETECTOR
Collimator to limit BSE and stray X-raysWDS DETECTORWindow usually made of beryllium(limited to sodium, atomic number 11) or thin plastic to facility detection down to boron (Atomic number 5)
Detector crystal silicon wafer with lithium added in. When hit by X-ray, creates a free electron and a hole. Keep at high voltage. For each 3.9eV from an X-ray, produce an electron and hole. This produces a pulse of current, the voltage of which is proportional to the X-ray energy.
must keep the crystal at LN temperature
1. prevent redistribution of lithiumField effect transistor (FET) amplifies signal from the crystal--2. reduce electronic noise
3. to keep crystal electrically isolated
Pulse processor: convert output of detector to voltage pulses,
the heights of the pulses are proportional to the X ray energywhen too many pulses come at one time, processor shuts off==>dead time
analog to digital converter (ADC) converts each pulse from processor to several pulses of equal height. Greater energy, more pulsesmulti channel analyzer (MCA ) sorts pulses into different channels of different energy
Diffractor--this is a crystal that reflects x-rays coming into the detector so they are positioned to hit the counter. Only those x-rays of the correct wavelength are correctly reflected to the counter. To do more than a few elements, you need several different crystals with different spacing in them that will reflect different wavelenght
Gas proportional counter (AT ROOM TEMP) --tube with wire in it
The X-rays that strike the counter ionize the contained gas and cause an increased current
FET Field effect transistor (FET) amplifies signal from the counter
Pulse processor: convers the output of detector to voltage pulses,
The heights of the pulses are proportional to the X ray energywhen too many pulses come at one time, processor shuts off==>dead time
SINGLE channel analyzer
OUTPUT
spectrum plot==>analysis of total area by elements
plot the x axis as energy, the y as counts of x-rays per second
dot maps ==> sums x rays of given window of energy or wavelength and portrays it on CRT. An element is chosen, and then each dot in the image is colored black if no x-rays from that element are detected for that spot, vs given a color if x-rays from that element are observed. By overlapping the maps of all the elements selected for observation, it is possible to ascertain the composition and distribution of elements in the surface (SEM) or in the slice(TEM) of the sample.
Qualitative analysis vs. quantitative analysis
qualitative analysis--what elements are present
quantitative analysis--what elements are present, and what is the concentration of each
quant is much more difficult
SET UP
1. accumulation time=> 30sec-10 minutes. Longer times give better statistics
2. beam current =>greater current, more X-rays. However, can overload
detector and get too much dead time (the detector shuts off and catches up in
the counting, then turns back on)
3.take off angle want to minimize absorption of x-rays by sample --usually 35 degrees off the horizontal (mount the detector on a tilt, or tilt the sample 35 degrees)
4 solid angle --want larger angle to collect more x-rays
ARTIFACTS
Systems peaks---created by interactions in the system
a) the beam can generate x rays from the aperture, which is thin and
allows the x-rays to come out the other side
use molybdenum aperture coated with tungsten and graphite stubs to reduce these effects
b) the X-rays can generate X-rays by fluorescence
very tough to control, use a collimator to block x-rays not coming from the sample
--sum peaks: two x rays arrive at almost the same time==>added together
collect with different dead time and look for changes in curves
--escape peaks: x-rays hit crystal and produce silicon x-rays instead
of electron hole pairs
the peak will therefore be reduced by that much energy
--secondary fluorescence: X-rays producing other x-rays of lower
energy
--specimen topography: optimum is flat surface
otherwise get absorption, variations in take off angle
--mass loss: can lose up to 30% over several minutes, lowering concentration of element or increasing apparent concentration of retained element
ZAF correction
A self absorption different elements making up sample will absorb more or less
F fluorescence X rays of one element produce x rays in lighter elements