Introduction to Electron Microscopy
1933 Ernst Ruska and Max Knoll first TEM
1938 von Ardenne first STEM
1942 V. Zworykin, J. Hillier and G. Snyder first SEM
1985 H Rohrer and G Binnig first scanning tunneling microscope
____________________________________________
Types of microscopes
Transmission Electron Microscope (TEM) (Fig.
1.1, 4.6) uses a wide beam of electrons passing through a thin sliced specimen
to form an image.
This microscope is analogous to a standard upright or inverted light microscope.
Stained areas of the sample absorb or scatter the beam, producing dark spots;
unstained areas appear light (Fig 1.3).
Scanning Transmission Electron Microscope (STEM)
uses a focused beam of electrons scanning through a thin sliced specimen to
form an image. The STEM looks like a TEM but produces images as does
an SEM (one spot at a time). The images are produced a spot at a time, as in
the scanning electron microscope, rather than all at one time as in the TEM.
It is most commonly used for elemental analysis of samples.
Scanning Electron Microscope (SEM) (Fig.
1.4, 5.3) uses focused beam of electrons scanning over the surface of thick
or thin
specimens.
This
microscope is analogous to the stereo or dissecting light microscope. Images
are produced one spot at a time in a grid-like raster pattern. Areas
reflecting lots of electrons appear bright; areas not relecting many electrons
appear dark (Fig 1.4).
Boston Museum of Science SEM animation: http://www.mos.org/sln/sem/sem.mov
Scanning Probe Microscopes (SPMs) are a family of instruments used for studying surface properties of materials from the atomic to the micron level. All these microscopes (atomic force, scanning tunneling microscope, etc.) use a probe tip that measures changes in surface characteristics as the tip scans over the surface of an object. Different kinds of images can be created depending on the nature of the probe tip. For example, in a scanning tunneling microscope, a fixed voltage is maintained between the tip of a probe and the conductive sample surface. If the surface dips, the voltage drops exponentially as the electron shells between probe tip and sample get further apart (tunneling current). The probe moves up or down to keep the voltage constant and thereby records the surface topography. The atomic force microscope uses a laser reflecting off the back side of the probe tip to determine the position of the probe tip. Ridges and valleys on the non-conductive sample cause the probe tip to rise and fall as the negative electron shells of the sample repel the negative electron shells of the probe tip (similar charges repel). These microscopes, although they may use interactions between electron shells, are not electron microscopes as used above.(Fig 5.29 -5.31)
Scanning probe microscopy of membrane bound protein complexes (from
http://www.ornl.gov/sci/GenomestoLife/areas/imaging.shtml)
atomic force microscope
| Diagram of atomic force microscope | Adapted from Hansma 2001. Ann. Rev. Phys. Chem 52:71-92 |
| . . The scanning probe microscope (SPM) is ideally suited for characterizations at or near the cell surface). The nanometer scale resolution of atomic force microscopy (AFM) can be used to image membrane bound proteins in a liquid environment and in many cases sub-molecular structure of individual proteins can be resolved. In addition to imaging, sensitivity of the AFM cantilever can be exploited to measure forces required to rupture bonds between biomolecules, within single protein molecules, and by tethering specific probe molecules to the cantilever to identify single molecule interactions with specific target molecules on surfaces. |
![]()
|
| Laser scanning confocal microscope (LSM) | http://www.confocal-microscopy.com/website/sc_llt.nsfl |
| uses an aperture to limit the images observed to a single focussed slice of a 3D fluorescent sample. Using a motor attached to the focus knob, a computer captures a series of these slices and reconstructs them into a 3D image. The aperture blocks a lot of light, so a high intensity source (a Laser) is necessary to provide sufficient illumination. The laser is focussed to a spot, and mirrors are used for Scanning the laser spot through the sample in a grid-like raster pattern. A pinhole aperture restricts the image collected by the detector to be the same focussed image observed in an optical slice of the sample (Confocal (same focus) images). All this is an accessory to a standard fluorescent Microscope. | ![]() |
Type of machines and illumination
Resolution is the ability to see two closely placed objects as discrete structures. It is the smallest distance one can see between two closely placed objects
Resolution in the light or electron transmission microscope:
under the best of conditions: human eye => 0.2mm
light scope => 0.2um
SEM => 3nm
TEM => 0.2nm
resolution is determined by the formula of Ernst
Abbe
resolving power= 0.61(wavelength)/ N.A. = 0.61(wavelength)/n
sin alpha
Substituting into the Abbe formula with everything else held constant, as the wavelength gets smaller, our resolving power gets better.wavelength: the peak to peak or trough to trough distance
radios => miles to mm infrared => mm to 800nm
light => 400-800nm
electrons => 20kV 0.008 nm
wavelength electron=1.23/acc.volt 1/2 80kV == 0.004 nm
A useful rule of thumb from the light microscope is that maximum theoretical resolution
l/2 wavelengthlight scope 550 nm/2 = 275 nm
UV scope 0.35nm/2 =175 nm wavelength of an electron will vary depending on accelerating voltage
actual limit of resolution in a field emission TEM is more like 0.1 nm due to the constraints of samples preparation and lens creation
electron =1.23/acc.volt1/2 1.23==>de Broglie's constant electron =1.23/80,0001/2 ==>1.23/282 ==>0.004nm
max resolution of microscope operating at 80kV is electron /2 ==>0.004/2=0.002
We can compare it to looking down from a cliff onto waves on the shore: when there are big waves, only big objects can be seen in the water
with smaller waves, smaller objects as well as big objects can be seen
--the smaller objects do not deflect the wave enough
Electron beam illumination/ interaction with the TEM specimenBeam of electrons passing through a thin slice interacts with sample in passing.
The ratio of transmitted and scattered electrons forms the image and sets contrast in the TEM
absorbed few electrons are completely absorbed--produce heat buildup
unscattered don't hit anything
inelastic interact with electrons of atoms and suffer small deflection, but large loss in energy
elastic interact with nuclei of atoms and undergo large deflection,but little or no loss of energy
Resolution in the scanning electron microscope |
|
Since the sample is illuminated with a spot that scans across the sample, the resolution of this microscope is determined by a number of different factors. Of prime importance is how small the spot is that exits the sample surface, and the magnification of the image for a given spot size. The wavelength formula is not applicable to this instrument. Resolution
is 3 nm tungsten, 1.5 nm field emission |
|
![]() |
![]() |
| Each spot in the SEM will show only an intensity of black/grey/white.
If the spot if larger than the detail of interest, the detail will be lost.
To see the stubs on this pollen grain, the spot size must be much much smaller than the small spont on the left. | |
Electron
beam illumination/ interactionwith
a SEM specimen |
|
This interaction region is pear shaped, referred to as the Monte Carlo configuration (interaction volume Fig 5.6-5.8) It will vary with accelerating voltage and atomic number |
![]() |
A
number of different interactions may take place within the sample
to produce signals that can be captured by the appropriate detector: |
|
| Auger e- | low energy, at surface, results from inner shell displacement |
| Secondary e- | low energy, mostly close to surface |
| Backscattered e- | high energy, deeper in sample |
| X-ray photons | deepest in sample |
| Absorbed e- | measure of conduction of electrons through sample |
| Cathodeluminescence | interactions causes sample to glow |